A Comparison Study of Transection Rate and Wound Surface Size Caused by a Manual and Motorized Sharp Punch in FUE Georgios Zontos MD, PhD Medical Director of Harklinikken, Denmark- University of Patras, School of Medicine, Patra, Greece


slide-1
SLIDE 1

A Comparison Study of Transection Rate and Wound Surface Size Caused by a Manual and Motorized Sharp Punch in FUE

Georgios Zontos MD, PhD

Medical Director of Harklinikken, Denmark- University of Patras, School of Medicine, Patra, Greece DISCLOSURES:

The Speaker has no relevant financial relationships or conflicts of interest to declare.

slide-2
SLIDE 2

Background

Some of the main disadvanges of FUE:

  • Time consuming
  • Transection rate (partial – total)

2

Georgios Zontos MD, PhD

  • White dots on the donor area

(hypopigmentation)

slide-3
SLIDE 3

Objectives

1) Follicle transection rate 2) The average value of wound surface size 3) The mean value of follicles obtained per graft 4) The harvesting rate 5) Ultimately to evaluate the quality of the obtained grafts in both manual and motorized punch technique

3

Georgios Zontos MD, PhD

slide-4
SLIDE 4

Materials and Methods

  • 23 patients
  • 0.9 mm sharp punch for both manual and motorized technique
  • Speed of rotation 15,000 – 17,000 rpm
  • Two equal surface size rectangles (8 cm

2 ), were marked on the donor area

  • Equal harvesting attempts made on each rectangle by using manual and motorized

technique

4

slide-5
SLIDE 5

Observation

Georgios Zontos MD, PhD

5 Manual punch Motorized punch

slide-6
SLIDE 6

Measurements

  • The follicle transection rate n =

𝑔𝑝𝑚𝑚𝑗𝑑𝑚𝑓𝑡 𝑢𝑠𝑏𝑜𝑡𝑓𝑑𝑢𝑓𝑒 𝑢𝑝𝑢𝑏𝑚 𝑜𝑣𝑛𝑐𝑓𝑠 𝑝𝑔 𝑔𝑝𝑚𝑚𝑗𝑑𝑚𝑓𝑡

x 100%

  • The value of λ = number of follicles / number of grafts for each case
  • The value of the extraction rate = number of grafts per minute in each case

6

Georgios Zontos MD, PhD Manual punch Motorized punch

  • The surface size of wounds in each rectangle was measured using a proprietary

image processing system

slide-7
SLIDE 7

Statistical analysis for both manual & motorized punch technique in 23 cases

  • All the data analyzed by applying independent sample t-test and it was found that p=0.000<0.05.

Because of this there is a statistically significant difference between the means with a 95% confidence interval of the difference. 7

Georgios Zontos MD, PhD

slide-8
SLIDE 8

Statistical analysis for both manual & motorized punch technique in 23 cases

8

Georgios Zontos MD, PhD

The transection rate using a motorized punch decreases by 77.94%.

0,00% 2,00% 4,00% 6,00% 8,00% 10,00% 12,00% transection rate

10,88% 2,40%

Transection Rate

Manual Motorized

slide-9
SLIDE 9

Statistical analysis for both manual & motorized punch technique in 23 cases

9

Georgios Zontos MD, PhD

The injury to the donor area using a motorized punch increases by 34.55%. This may be the resultant effect of the higher rotational motion of the punch and the nutation of the punch cylinder axis.

0,00 0,20 0,40 0,60 0,80 1,00 1,20 1,40 1,60 Wound Surface (mm2)

1,10 1,48

Wound Surface

Manual Motorized

slide-10
SLIDE 10

Statistical analysis for both manual & motorized punch technique in 23 cases

10

Georgios Zontos MD, PhD

The ratio λ using a motorized punch increases by 12.77%.

0,00 0,50 1,00 1,50 2,00 2,50 3,00 3,50 Ratio λ

2,82 3,18

Ratio λ

Manual Motorized

slide-11
SLIDE 11

Statistical analysis for both manual & motorized punch technique in 23 cases

11

Georgios Zontos MD, PhD

The extraction rate using a motorized punch increases by 37.12%. This in combination with the simultaneous decrease in the transection rate makes the harvesting process more efficient because more hairs are collected using the minimum harvesting attempts in a shorter period of time.

0,00 2,00 4,00 6,00 8,00 10,00 12,00 14,00 Extraction Rate (Grafts/min)

9,59 13,15

Extraction Rate

Manual Motorized

slide-12
SLIDE 12

Evaluation of the quality of the obtained grafts

12

The motorized technique seems to be more efficient in producing better quality of grafts. Despite being generated by the same punch size ,the grafts produced by the motorized punch appear to possess more protective tissue than those generated by the manual technique

Manual punch Motorized punch

slide-13
SLIDE 13

Conclusion

13

Georgios Zontos MD, PhD

The motorized punch makes the FUE technique quicker and more efficient producing a larger number and better quality of grafts. However, the larger wound trauma to the donor area increases the probability of noticeable scaring within the donor area.